• Anesthesia Product Specifications

    Theory of Operations  

    Tapsystem Model 2A Device Specifications

    Model 550 TAPSCOPE

    EC-View ECG Adaptor

    2A Pole Mount

     

     


     

    Theory of Operations:

    The Model 2A generates cardiac stimulation pulses which are administered to the patient's heart via
    the TAPSCOPE 550, an 18 Fr esophageal stethoscope with bipolar pacing electrodes. All circuitry is contained on a single printed circuit board mounted behind the front control panel. When the power switch is toggled to ON, DC power from 2 D-cell batteries is applied to a power regulation circuitry, a DC-DC converter, and a rectangular pulse forming circuit with a constant current output stage whose amplitude can be varied over a broad range of load impedances. Hospital Biomedical Engineers requiring a complete Model 2A Service Manual (including circuitry schematics and parts list) should contact CardioCommand Customer Service at 1-800-231-6370.

    Back

     



    Device Specifications:

    Electrical

    RATE:              Continuously variable, 50-200 BPM
    AMPLITUDE:    Continuously variable, 5.5-40 mA rectangular waveform
    DURATION:      Constant at 10 msecs/pulse
    Accuracy:        Stimulus parameters within ± 10 % of full scale
    On/Off:             Toggle switch
    Impedance:      Constant current over loads up to 2k ohms at 20 mA
    Defibrillation:     Protected up to 400 Joules

    Physical   Size:              2.92" (7.4 cm) high x 4.73" (12) deep x 6.70" (17) wide
    Weight:         0.7 lbs (.32 kg) without batteries
    Material:       Impact-resistant ABS plastic enclosure
    Power Supply Battery type: 1.4 V Alkaline D-cell (2). The unit should operate normally for 70 hours with fully charged batteries at specified operating conditions.
    Continued operation on low-batteries may result in erratic output.
    Warning: Use of batteries other than those recommended by manufacturer may result in the Model 2A not operating within specifications.
    Cable 10 foot cable assembly provides connection between TAPSCOPE
    esophageal cardiac pacing/recording catheter and Model 2A
    TAPSYSTEM stimulator.
    Indicator Lights A yellow LED illuminates to indicate low battery.
    A green LED indicator flashes with each stimulus pulse delivered only
    when an appropriate load impedance is connected to the stimulator.

    Back


     

    Model 550 TAPSCOPE:

    18 French esophageal stethoscope with MA100 Series Thermistor and bipolar electrodes for connection to Model 2A Stimulator (cardiac pacing) or to EC-View Adaptor (ECG recording). Esophageal catheter has markings indicating Depth of Insertion of electrodes from the lips (cm).
    Each single-use, disposable TAPSCOPE is individually pouched. 25 units are shipped per carton.

    Back


     

    EC-View ECG Adaptor:

    When the TAPSCOPE is not used for pacing, the bipolar electrodes can be used for monitoring
    of the esophageal ECG using the EC-View Adaptor. Two snap contacts allow attachment of
    standard ECG leads for connection with ECG monitors or other recording devices. Adaptors
    are individually pouched and shipped as single units.

    Back

    2A Pole Mount:

    Metal frame for attaching Model 2A TAPSYSTEM to hospital pole.

    Back